![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086441.webp)
18-60
SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
SENSe1 subsystem
The Sense1 subsystem is used to configure and control the measurement functions of the
SourceMeter. Many of the commands are global, where a single command affects all func-
tions. Some commands are unique to a specific function. For example, you can program a
unique range setting for each basic function (amps, volts, and ohms).
A function does not have to be selected before you can program its various configurations.
Whenever a programmed function is selected, it assumes the programmed states.
The commands for this subsystem are summarized in
Table 18-6
.
Select measurement functions
CONCurrent <b>
[:SENSe[1]]:FUNCtion:CONCurrent <b>
Control concurrent measurements
Parameters
<b> =
0 or OFF
Disable concurrent measurements
1 or ON
Enable concurrent measurements
Query
:CONCurrent?
Query state of concurrent measurements
Description
This command is used to enable or disable the ability of the instrument
to measure more than one function simultaneously. When enabled, the
instrument will measure the functions that are selected. (See
“[:ON]
<function list>,” page 18-61
,
“OFF <function list>,” page 18-61
, and
“ALL,” page 18-62
.)
When disabled, only one measurement function can be enabled. When
making the transition from :CONCurrent ON to :CONCurrent OFF, the
voltage (VOLT:DC) measurement function will be selected. All other
measurement functions will be disabled. Use the :FUNCTion[:ON]
command to select one of the other measurement functions.
NOTE
For the Model 2430 Pulse Mode, concurrent measurements are always disabled.
Sending this command results in error +831.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...