11-2
Triggering
2400 Series SourceMeter
®
User’s Manual
Front panel trigger operation
Front panel trigger model
NOTE
For the Model 2430, the following discussion on the trigger model pertains
explicitly to the DC Mode of operation. The trigger model operates a little dif-
ferently for the Pulse Mode. These differences are covered in “
Pulse Mode trig-
gering (Model 2430)
” later in this section.
The flowchart in
Figure 11-1
summarizes triggering for front panel operation. The trigger
model is modeled after the remote commands used to control triggering. (See
“Remote
trigger commands,” page 11-18
.) Key trigger model settings are included in the flowchart.
Note that the BENCH defaults are denoted by the “
✛
” symbol.
The primary actions of the trigger model are Source, Delay, and Measure. The source
action outputs the programmed voltage or current value, and the programmed delay pro-
vides a settling period for the source before the measurement is performed.
The trigger model consists of two layers (Arm Layer and Trigger Layer) to provide versa-
tility. Programmable counters allow operations to be repeated, and various input and out-
put trigger options are available to provide source-measure synchronization between the
SourceMeter and other instruments (via the Trigger Link).
Unless otherwise noted, the programmable aspects of the trigger model are performed
from the CONFIGURE TRIGGER menu. (See
“Configuring triggering,” page 11-8
.)
Event detection
In general, operation is held up at an Event Detector until the programmed event occurs.
Note however, that if an event detector has a bypass, operation can be programmed to loop
around the event detector.
Idle
The SourceMeter is in idle when it is not operating in the Arm Layer or Trigger Layer of
the trigger model. When in idle, the ARM annunciator is off. To take the SourceMeter out
of idle, turn the output ON.
The SourceMeter can be returned to idle at any time by selecting the HALT menu item of
the CONFIGURE TRIGGER menu. (See
“Configuring triggering,” page 11-8
.)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...