8218 pH pen manual

 

 

 

1. Introduction 

8218  pH  pen  is  designed  with  waterproof.  It  is  an  ideal  instrument  in  pH  and  the  temperature 

measurement for water in family fishbowl and university lecture experiment. The meter has functions in 

auto calibration, auto temperature compensation and 6 minutes auto power off.     

     

2. Specification 

2.1 Technical parameter 

pH

range: 0.00

14.00 

Resolution: 0.01 

Accuracy: 0.05 

Temperature: range

0

60.0

 

Resolution: 0.1

 

Accuracy: 0.5

 

2.2 Environment temperature: 5

35

 

2.3 Power

A76 button cell, 1.5V×4 ea 

2.4 Dimension

175×22×38mm 

2.5 Net weight

85g 

   

3. Usage and maintenance 

3.1 Measurement: remove the protection cap from the electrode, press [ON/OFF] to turn on the power. 

Put the electrode into the solution to be tested, then the meter shows the pH and temperature data of 

the solution. After measurement, clean the electrode with water, put on the protection cap, press and 

hold on [ON/OFF] to turn off the meter.     

3.2 The meter has got calibrated before its shipment. For calibration operation, please refer to chapter “4. 

Calibration” of this manual   

3.3 Measurement after long time not used (over two months), the glass bulb of the electrode should be 

dipped in the water for 30 minutes, which should be under water 2-3 cm.     

3.4 Maintenance: the electrode should be well cleaned with water. If the meter displays a cell graphic, it 

means  the  cell  voltage  is  low,  please  replace  it  with  a  new  cell.  Please  pay  attention  to  the  cell 

polarity during the replacement.       

 

4. Calibration 

4.1 Press [CAL] button to enter into calibration status, 

CAL

is then twinkling on the screen. Please 

follow up the steps shown on the screen and put the electrode into a corresponding standard solution. 