05.13 / V1.0
4-5
UNIDRIVE
®
S III
D
Connect the footswitch.
✓
The error message
E12
(no motor) is displayed.
D
Connect a handpiece to the motor socket M1.
✓
The handpiece is recognized and the corresponding information displayed. e.g. P0, P1, etc.
D
Press the right footswitch pedal.
✓
The motor starts.
D
Release the footswitch pedal.
✓
The motor stops immediately.
D
Change the direction of rotation by pressing the mid button of the footswitch.
✓
The displayed direction of rotation changes.
For devices with motor socket M2:
D
Connect a handpiece to the motor socket M2.
D
Select the handpiece M2 by pressing the M button.
✓
The handpiece is recognized and the corresponding information displayed.
D
Press the right footswitch pedal.
✓
The motor starts.
D
Release the footswitch pedal.
✓
The motor stops immediately.
D
Change the direction of rotation by pressing the mid button of the footswitch.
✓
The displayed direction of rotation changes.
For devices providing a pump:
D
Press the left footswitch pedal.
✓
The pump starts running.
D
Change the flow of the pump pushing the
+/-
buttons on the display.
✓
The pump has to run regularly at all settings.
D
Open the tensioning lever of the pump.
✓
The error message
E16
(irrigation pump not closed
)
is displayed.
D
Close the tensioning lever of the pump again.
✓
The error message disappears.
Содержание 20 7010 20-1
Страница 4: ......
Страница 8: ......
Страница 9: ...I UNIDRIVE S III ENT UNIDRIVE S III ENT UNIDRIVE S III ENT 40701620 1 40701620 1 40701620 1 ...
Страница 10: ......
Страница 12: ......
Страница 127: ......
Страница 129: ......
Страница 130: ......
Страница 131: ......
Страница 134: ......
Страница 146: ......
Страница 148: ......
Страница 158: ......
Страница 166: ......
Страница 168: ......
Страница 169: ......