107
Table 206
Recommended separation distances between portable and mobile RF communications equipment
and the UNIDRIVE
®
S III ENT model 40 7016 20-1
The UNIDRIVE
®
S III ENT model
40
70
16
20-1 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or user of the unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF commu-
nications equipment (transmitters) and the unit as recommended below, according to the maximum output power of the communications equipment.
Separation distance d [m]
according to frequency of transmitter
Rated maximum output power
of transmitter
[W]
150 kHz to < 80 MHz
D
80 MHz to < 800 MHz
D
800 MHz to 2.5 GHz
D
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters [m] can be estimated using the equa-
tion applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts [W] according to the transmitter
manufacturer.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
Anexo
Appendix
Anhang
Содержание 20 7010 20-1
Страница 4: ......
Страница 8: ......
Страница 9: ...I UNIDRIVE S III ENT UNIDRIVE S III ENT UNIDRIVE S III ENT 40701620 1 40701620 1 40701620 1 ...
Страница 10: ......
Страница 12: ......
Страница 127: ......
Страница 129: ......
Страница 130: ......
Страница 131: ......
Страница 134: ......
Страница 146: ......
Страница 148: ......
Страница 158: ......
Страница 166: ......
Страница 168: ......
Страница 169: ......