3-7
Starting the System
After an E-Stop Emergency Shutdown
To recover from an E-Stop, the E-Stop button that was pushed needs to be reset. Pull
its red knob out and turn it clockwise until it clicks into position.
Figure 24: Resetting E-Stop button
To start after an E-Stop emergency shutdown:
1. Make sure all switches are ready for power-up:
2. Press the green
POWER ON
button on the front of the workstation. The computer
and internal electronic systems start up.
Figure 25: SYSTEM POWER - ON/OFF buttons
3. The laser’s chiller will then auto-start and load its run-mode program.
4. Log on to Windows with user name
Operator
. The factory-set password is
1
.
5. Start the software by double-clicking their icons on the desktop.
6. In the User Access Control dialog box, log in with the assigned user name and
password. Factory settings let you log in using
Operator
, password
1
.
Switch
Location
Setting
Laser power switch
Back of laser power supply
Off
Laser key switch
Front of laser power supply
“0” Position
Emergency Stop buttons
Front panel and inside back door
Pulled out
1
2
Green Power ON Button