10-5
Processing Area Ventilation
Debris Collection Filter (optional)
An optional debris collection HEPA filter may be required. If so, it will need to be
connected to the processing area ventilation via the 1-inch male NPT fitting on the
roof of the enclosure. . The unit is on casters and can be placed anywhere, however it
will need to be situated within 10 ft (3m) of the enclosure to minimize the length of the
suction hose and maximize the efficiency of the filter.
A debris collection filter with a ducted output is also available. Contact IPG
Microsystems Customer Service for details.
Specifications
❑
Hose Inlet Connection: 100mm
❑
Airflow: 380m3/hr
❑
Pressure: 96mbar
❑
Electrical Data: 115-230v 1ph 50/60Hz Full load current: 12.5 amps / 1.1kw
❑
Dimensions: 37.5 cm wide
37.5 cm deep
59 cm high (14.8 in
14.8 in
23.2 in)
❑
Weight: 40 kg (88 lb)
Figure 89: Typical Debris Collection HEPA Filter
100mm Hose Inlet Connection