8 - 2
Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00
8 - Troubleshooting
Troubleshooting Table
Front safety panel does not
lock properly
Mechanical failure of the
door locks
Switch off the instrument. Contact
your local service organization
Problem, error on liquid handling arm (LiHa) and tips
Positioning error
X, Y or Z-drive blocked
Crash
Hardware defective
Check for obstacles
Check container, rack and carrier
positions
Contact your local service organiza-
tion
See
Problem, error on Positive Identification, PosID
Positioning error
Hardware defect
Contact your local service organiza-
tion
Barcode not read
Barcode label not facing
barcode scanner
Check container position on carrier
See
Bad barcode label quality
Check with new barcode label
See
Barcode type not accord-
ing to specifications
Check if barcode type is permissible
See
Barcode label position not
according to specifications
Check barcode label position on
container
See
Barcode type not specified
in software
Check the settings in the application
software
Laser output window dirty
Clean output window
See
Alignment barcode on bar-
code flag not read
PosID adjustment/setup not
correct
Contact your local service organiza-
tion
Carrier or tube presence
not detected
“No Tube” sensor dirty
Clean “No Tube” sensor
See
Unusual noise during
movement
Worn out or damaged parts
Contact your local service organiza-
tion
Problem, error on robotic manipulator arm, RoMa Standard
Microplate not picked up
No microplate on carrier
Cannot pick up microplate
Put microplate on carrier
Set gripper position
Clean grippers
Unusual noise during arm
movement
Worn or damaged parts
Contact your local service organiza-
tion
Tab. 8-1
Troubleshooting table (cont.)
Problem, error
Possible cause
Corrective measures