6 - 16
Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00
6 - Operation
Operating in Routine Operation Mode
Wash Station/Waste Tubing
Installing the
Waste Tubing
When installing the waste tubing, pay attention to the following:
ATTENTION
Liquid spillage on the worktable.
To prevent overflow of the wash station the waste tubing must be routed in such a
way that the back pressure is as low as possible.
The waste tubing must not be longer than necessary.
The waste tubing must not be kinked or squeezed (reduction of clear cross
section).
The waste tubing must not rise after wash station (back pressure).
The waste tubing must not sag (back pressure).
The lower end of the waste tubing must not be in the liquid (back pressure).
Fig. 6-5
Correct and disadvantageous course of waste tubing
A
A
B
B
D
C
C
E
F
G
Correct installation of waste tubing
A
Wash station
B
Worktable
C
Waste tubing
D
Correct course of waste tubing
Incorrect installation of waste tubing
E
Rising waste tubing
F
Sagging waste tubing
G
Waste tubing reaching into liquid