6 - Operation
Operating in Routine Operation Mode
ATTENTION
Incorrect identification of the carrier (strip rack).
The carrier barcode is associated with the corresponding tube size. Therefore, the
carriers are not handled correctly if the inserts are exchanged.
Do not change the strip rack inserts.
Do not exchange the carrier barcode flags.
Note:
The filling level of tubes, troughs and containers must not exceed 80% to
avoid spillage during PosID reading.
Tab. 6-2
Minimum inner diameter for primary sample tubes
Tip type
Tube diameter, inside
Fixed tips
7 mm