3 - Technical Data
Technical Data
3.2.4 Environmental Conditions
All instruments are intended for indoor operation and storage only. The tables
below give an overview.
ATTENTION
Barcodes cannot be read due to the influence of sunlight or other light sources on
the barcode scanner.
Do not expose the instrument to direct sunlight.
Do not install strong light sources that may impair the function of the barcode
scanner near the instrument.
Operating
Conditions
Pipetting
Conditions
Storage
Conditions
Transport
Conditions
Operating temperature
15°C to 32°C (59°F to 90°F)
Operating humidity
30% to 80% relative (non condensing)
at 30°C (86°F) or below
Operating altitude
max. 2000 m above sea level
Pipetting temperature
20°C to 27°C (68°F to 80.6°F)
Pipetting humidity
30% to 60% relative (non condensing)
at 25°C (77°F) or below
Storage temperature
1°C to 60°C (34°F to 140°F)
Storage humidity
5% to 80% relative (non condensing)
at 30°C (86°F) or below
Transport temperature
-20°C to 60°C (-4°F to 140°F) for maximum 24 hours
Transport humidity
20% to 80% relative (non condensing) for maximum 24
hours