7 - 20
Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00
7 - Preventive Maintenance and Repairs
Maintenance Tasks
“No Tube”
Sensor
To clean the “No Tube” sensor, proceed as follows:
1
Switch the instrument off.
2
Remove the carriers in front of the PosID to gain access to the PosID.
Fig. 7-4
PosID barcode scanner
2
Switch the instrument off.
3
Remove the carriers in front of
the PosID to gain access to the
PosID.
4
Visually check the laser output
window (B) for cleanliness.
5
Moisten a lint-free tissue with
alcohol and clean the output
window, if necessary.
A
B
Fig. 7-5
PosID “No Tube” sensor
3
Slide the PosID gripper (A)
back to gain access to the “No
Tube” sensor (B).
4
Moisten a lint-free tissue with
alcohol and clean the front
surface of the “No Tube”
sensor.
A
B