7 - Preventive Maintenance and Repairs
Maintenance Schedule
Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00
7 - 7
At End of Day
Tab. 7-7
Daily maintenance at end of day in chronological order
Instrument/
Component
Maintenance Task
Reference
Tips
Clean inside and outside
See section
Clean standard tips
Caustic soda solution (1% NaOH)
Check all tubing, tubing connec-
tions, syringes
Carriers and racks
Clean using a detergent or anti-
septic solution
See section
Worktable
Clean
See section
Safety panel
Clean
See section
Wash station
Clean using a detergent or anti-
septic solution
System liquid container
Rinse with water and fill up
Waste container
Clean using a detergent or anti-
septic solution
Waste tubing
Clean using a detergent or anti-
septic solution
RoMa standard
Clean gripper fingers using alco-
hol or acetone
–
Plate washer
Leave filled with deionized water
overnight
–
Liquid system
Check for leakages after every 8
hours of operation
If liquids other than water are
used as system liquid, flush with
deionized water