7 - Preventive Maintenance and Repairs
Maintenance Tasks
Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00
7 - 17
7.3.5 Worktable
WARNING
Possible worktable damage
Only clean the worktable with small amounts of cleaning agent, e.g. with a
damped cloth.
Do not spill cleaning agent on the worktable.
Cleaning the
Worktable
Perform the following procedure to clean the pipetting instrument’s worktable:
1
Remove all racks and carriers from the worktable.
2
Wipe the surface of the worktable with a suitable cleaning agent (e.g. alcohol,
disinfectant) to remove any spilled reagent.
3
If necessary, additionally clean with water.
7.3.6 Safety Panels
Cleaning the
Safety Panels
Perform the following procedure to clean the safety panels.
Wipe the inner and outer surface of the safety panels with a suitable cleaning
agent, e.g. water, alcohol or disinfectant, to remove any spilled reagent or
sample.
If necessary, additionally clean the surface with water or alcohol.
7.3.7 Liquid Containers
System Liquid
Container
To prevent deposition of crystals and growth of micro-organisms in liquid
containers, clean all liquid containers at least once a week. Make sure to allow
solvents (e.g. ethanol) to evaporate before filling reagents into the containers
again.
Waste
Container
Clean the waste container at least once a day.
WARNING
Contamination through waste liquid, if the containers are installed wrongly.
Make sure not to mix up the system liquid container and the waste container.