7 - 16
Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00
7 - Preventive Maintenance and Repairs
Maintenance Tasks
7.3.4 Wash Station
Note:
Always make sure that the wash station is installed in the correct grid
position when it has been removed. If the grid position has changed, verify the
corresponding definitions in the application software.
7.3.4.1 Cleaning the (Standard) Wash Station
Cross
References
List of cross references to information provided in other sections:
The wash station can come in contact with reagents and samples. If a spillage
occurred, the wash station needs to be removed from the worktable for cleaning.
Clean the wash station as follows:
1
Wipe the surface of the wash station with a suitable cleaning agent (e.g.
water, alcohol, disinfectant) to remove any spilled reagent.
Note:
Do not use bleach to clean the wash station and do not clean it in a
laboratory washing machine.
2
If necessary, rinse the wash station and clean it additionally with water or
alcohol.
4
Clean the wash station as described above.
5
Clean the worktable.
Refer to cross references above.
6
Reinstall the wash station on the worktable.
Make sure that the wash station is pushed all the way back to the stop during
installation.
Subject
Reference
Clean worktable
See section
Fig. 7-3
Wash station
If necessary, remove the wash
station from the worktable.
1
Open the front access panel
(C).
2
Loosen the nut (B).
3
Pull the wash station (A) to
the front (see arrow).
A
B
C