3 - 8
Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00
3 - Technical Data
Requirements
3.3 Requirements
3.3.1 Computer Requirements
USB port (standard)
RS232 port (optional)
3.3.2 Software Requirements
This instrument is compatible with IAC 6.0 or higher.
3.3.3 System Liquid Requirements
System Liquid
System liquid refers to a liquid which fills the liquid system and is used as wash
fluid.
Standard liquid
–
Deionized or distilled water with a conductivity between 0.5 µS/cm and 10
µS/cm
The system liquid must be free of particles.
Make sure that the system liquid container is clean.
The system liquid must be free of air bubbles and must be room temperature.
To reach the pipetting performance we recommend degassing the system
liquid. For further information on this issue, please contact your responsible
application specialist.
In order to ensure that during operation no air bubbles form in the pipetting
tubing, a sufficient quantity of system liquid must circulate in the system. We
recommend at least 60 ml per hour.
Any additives to the system liquid must be validated to evaluate the influence on
the pipetting performance and the overall analytical process.
3.3.4 Sample Requirements
Cross
References
List of cross references to information provided in other sections:
The instrument is validated for pipetting deionized water. Other liquids are only
allowed after validation according to laboratory practice and state-of -the art by
the kit manufacturer or operator of the system.
For sample preparation refer to cross references above.
Subject
Reference
Sample preparation
See section