7 - 8
Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00
7 - Preventive Maintenance and Repairs
Maintenance Schedule
7.2.3 Maintenance Table: Weekly Maintenance
Weekly
Maintenance
Note:
The weekly maintenance should be performed on the last working day of
each week.
7.2.4 Maintenance Table: Yearly Maintenance
Every twelve
Months
Tab. 7-8
Weekly maintenance
Instrument/Component
Maintenance Task
Reference
Liquid system
Clean
See section
System liquid container
Empty and clean
See section
Waste container
Empty and clean
See section
Liquid handling arm,
Robotic manipulator arm
Clean front arm guide
See section
PosID
Clean laser output window and
“No Tube” sensor
See section
Clean PosID working area of
the worktable (abrasion)
Lint-free cloth and alcohol
Tab. 7-9
Yearly maintenance
Instrument/Component
Maintenance Task
Reference
LiHa
Liquid handling performance verifi-
cation testing with QC Kit
(optional)
See section
dling Performance Verification
Testing”,
Complete Infinium LiHa
and Infinium RoMa
Clean system
Call Illumina customer support
to perform the task.
Frontal arm guide
Clean
Call Illumina customer support
to perform the task.
Worktable
Visually inspect worktable grids for
wear and replace
if necessary.
Call Illumina customer support
to perform the task.
LiHa
Visually inspect moving parts,
especially Y-belt, for wear and
replace faulty parts.
Check parts for abrasion; wipe
wear debris off, if necessary.
Call Illumina customer support
to perform the task.