Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00
6 - Operation
Operating in Routine Operation Mode
Safe Worktable Layout
ATTENTION
Unsafe layout of the worktable can cause e.g.:
Loss or dropping of microplates
Spillage of hazardous liquids because of collisions or too high filling level
(more than 80%) of cavities
Spillage due to imprecise pipetting in 96-well microplates placed on the Te-
Link
Cross-contamination because critical elements are placed near the wash
station waste (splashes).
Before and during instrument use, check the worktable for the safety of its layout.
Liquid System / Liquids
ATTENTION
Leakage of the liquid system.
By the continuous up and down movements of the syringes during operation, the
syringe and plunger lock screws may get loose, if not tightened properly. This
causes leakage of the liquid system.
Check the plunger lock screws and syringe screws and tighten manually
before switching the Infinium LiHa and Infinium RoMa on.
ATTENTION
To ensure a proper liquid flow, make sure that the tubings are not twisted or
inhibited from free flow.
ATTENTION
Instruments are intended for indoor operation with controlled temperatures.It is
important to maintain constant temperature plus air gaps.