6 - Operation
Operating in Routine Operation Mode
Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00
6 - 11
6.3.4.1 Carriers
Cross
References
List of cross references to information provided in other sections:
Carrier
Positioning
Slide carriers over the positioning pins until they abut on the lock pins.
Make sure that the barcode on the carrier corresponds to the settings in the
Application Software.
Carrier Fixation
and
Replacement
The positioning pins hold the carriers in defined positions, but still allow carriers to
be exchanged during an application. A rail in the carrier base fixes the carrier in X-
direction, the stop pins in the third row on the worktable fix the carrier in Y-
direction. When prompted to do so by the software, the operator can replace a
carrier during an application.
ATTENTION
Make sure that the stop pins limit the carrier movement correctly, otherwise
crashes or incorrect pipetting may occur.
Positioning
Pins
If a positioning pin is damaged, replace it immediately.
Refer to cross references above.
Place the carriers only on the provided positions as the instrument is adjusted to
these positions. Carriers placed e.g. on the left of positioning pin 1 can cause
mechanical problems (collision) or errors in the identification of barcoded
samples.
Placing Carriers
All carriers must be in close contact with the worktable, so that the capacitive
liquid level detection is guaranteed. For this purpose, clean the carriers and the
worktable in regular intervals.
Refer to cross references above.
Make sure that the correct rack is used for the carrier.
If a carrier is damaged, replace it immediately.
Carrier ID
Each carrier ID must be unique.
Subject
Reference
Carrier cleaning
See section