3.4 Setting Measurement Configuration
67
3
Chapte
r 3 Me
asu
remen
t Proced
ure
The sampling rate is automatically set to 1/100 of the selected time axis range.
If MR8990 Digital Voltmeter Unit is installed, the sampling rate for the chan-
nels of that unit is set to 1/50.
Example: When 8966 is installed on unit 1 (channels 1 and 2) and MR8990 on
unit 2 (channels 3 and 4), and the time axis is set to 1 s/div
Sampling rate of unit 1 (8966): 10 ms
Sampling rate of unit 2 (MR8990): 20 ms
_______________________________________________________________
• The timebase and sampling rate can be set independently. The sampling rate
is selected depending on the timebase setting.
• When the following timebase values are selected, displayed waveforms are
compressed in the horizontal (time axis) direction as shown.
20 ms/div
→
x1/2
10 ms/div
→
x1/5
_______________________________________________________________
Common
To minimize noise during measurement
If the sampling rate is set too fast, when the input waveform ampli-
tude is small, the difference between maximum and minimum values
may become quite large as a result of sudden impulses such as
noise. To prevent such phenomena, select a slower sampling rate or
enable the input module's lowpass filter (
The data refresh rate is not allowed to exceed the maximum sampling rate of the
input module.
During the period when data are not updated, the same data are measured,
resulting in a stair-shaped waveform. Even when the same signal is input simul-
taneously by two units, the data may differ due to differences in sampling rate,
frequency bandwidth, and frequency response.
Data refresh rate for various units
Module
Max. time axis range
Max. sampling setting
See
8966
5
μ
s/div
50 ns (20 MS/s)
-
8967
Dependent on data re-
fresh setting
Dependent on data re-
fresh setting
See:
8968
100
μ
s/div
1
μ
s (1 MS/s)
-
8969
U8969
500
μ
s/div
5
μ
s (200 ks/s)
-
8970
Dependent on setting Dependent on setting
See:
8971
100
μ
s/div
1
μ
s (1 MS/s)
See:
8972
Dependent on re-
sponse setting
Dependent on re-
sponse setting
See:
8973
5
μ
s/div
50 ns (20 MS/s)
-
MR8990 Dependent on NPLC
setting
Dependent on NPLC
setting
See:
U8974
Dependent on NPLC
setting
Dependent on NPLC
setting
See:
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......