7.9 Making Detailed Settings for Input Modules
176
Channels installed with U8793 cannot be measured.
See:
"Opening the [Each Ch] sheet, Making a Channel Selection" (p.159)
Type
Selects the waveform type.
Select
Refer the instruction manuals of Model U8793, MR8790, and MR8791 for the
settings at the time of program selection.
Frequency
Sets the frequency of output signal.
Amplitude
Sets the amplitude of output signal.
The output voltage with guaranteed accuracy is the sum of the amplitude and the
offset, between -10 V and +10 V. If the sum of the amplitude and the offset is set
outside the guaranteed accuracy range, parts of the waveform will be clamped to
the upper limit, approxi16 V and the lower limit, approximately -11 V.
7.9.12 U8793 Arbitrary Waveform Generator Unit Set-
tings
Selections
Description
DC
DC output (Default setting)
Sine
Sine wave output
Square
Rectangular wave output
Pulse
Pulse wave output
Triangle
Triangular wave output
Ramp-up
Ramp-up wave output
Ramp-down
Ramp-up wave output
Arbitrary
Outputs the created waveform
Program
Outputs the waveform set in the program
0 Hz to 100000 Hz
0.000 V p-p to 20.000 V p-p
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......