13.1 Waveform GO/NG Evaluation (MEM, FFT Function)
292
• Trigger mode: SINGLE Measurement continues until stop mode conditions
are fulfilled and then stops.
• Trigger mode REPEAT, AUTO Recording and waveform evaluation is carried
out continuously. Click
[STOP]
to terminate the measurement.
• When "auto save" is set to ON, data are stored on media when the stop condi-
tions are satisfied.
• When memory segmentation is ON, data is recorded to the memory block only
when the stop conditions are satisfied.
• During waveform evaluation, evaluation is performed concurrently with mea-
surement when the time axis range is 1 ms/div or less with the roll mode set-
ting ON or during automatic operation with the display at 100 ms/div or less as
a result of the time axis magnification factor. At all other times, evaluation is
performed after measurement is complete.
• When performing evaluation after measurement is complete, this sequence of
two alternating operations (capturing data and performing the evaluation) is
performed repeatedly, so that data is not captured while the evaluation is in
progress. Therefore data are not captured while the evaluation is in progress,
which means that the input signal is not being continuously monitored. The
time required for evaluation is on the order of 100 ms.
• When performing evaluation and measurement concurrently, the waveform is
displayed and evaluated while measurement continues. However, please note
that processing is not necessarily conducted in real time.
• If a high setting is chosen for recording length or if compression is used, the
evaluation cycle becomes slower.
• Waveform scrolling is not available while waveform evaluation is ON. To use
waveform scrolling, you should first turn waveform evaluation OFF.
• Although the waveform display will change if you change display-related set-
tings after measurement completes, the waveform evaluation results from
when the measurement was made are retained. To repeat the evaluation pro-
cess for the waveform after the new settings have been applied, perform
waveform evaluation on the Status basic settings screen.
See:
"6.5.1 Magnifying and Compressing Horizontal Axis (Time Axis)" (p.127)
• When using the highlight function with the FFT function, only the highlighted
data is evaluated.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......