18.3 Error Messages
370
32
Zero-adjustment needed.
Perform zero-adjustment.
"2.6 Adjusting the Zero Po-
sition (Zero-Adjust)" (p.57)
33
Disabled key.
Close the dialog.
-
34
Invalid key pressed (Overlay)
The key operation is prohibited because Over-
lay is enabled.
"7.3 Displaying New Wave-
forms Over Past Waveforms
(Overlay)" (p.144)
36
No trigger has been set.
Set trigger criteria.
"Chapter 8 Trigger Set-
tings" (p.185)
38
Use of logic channels has reduced ana-
log waveform resolution from 16 to 12
bits.
Using channels LA - LB reduces resolution of
Ch1 - Ch2 to 12 bits.
"7.9 Making Detailed Set-
tings for Input Modules"
(p.159)
39
Auto balance failed.
Check whether a strain gauge transducer is in
an uncharged state, and that it is connected
correctly.
"7.9.4 Settings for the 8969
and U8969 Strain Unit"
(p.163)
40
Voltage Sag triggering is disabled. (Val-
id time base range: 20
μ
s/div to 50 ms/
div)
Voltage Sag triggering can be used only when
the time base is between
20
μ
s/div to 50 ms/div
"8.3.1 Analog Trigger Set-
tings and Types" (p.188)
42
There is not enough data in the memo-
ry.
Measure the data portion required for calcula-
tion.
"10.1 Waveform Calcula-
tion Workflow" (p.224)
43
Stopped.
-
-
46
Channels may no longer be able to be
used by frequency units due to logic
use.
If logic channels A and B of the machine are
used, the respective frequency units at chan-
nels 1 and 2 cannot be used.
47
:This unit cannot be used because it is
not AAF aligned.
Display the
[Init]
sheet on the SYSTEM
screen.
Execute the
[System Information]
and check
the software items.
If
[Not AAF aligned]
is displayed, send the
unit for repairs.
-
48
There are units on which AAF cannot be
set to On because of lack of adjustment.
50
Roll Mode is not available.
The Roll Mode cannot be used when Overlay
is enabled.
"7.2 Displaying Waveforms
During Recording (Roll
Mode)" (p.143)
51
Cannot use. (Pre Trigger)
The pre-trigger function is not available when
using external sampling.
"Chapter 16 External Con-
trol (MR8741 Only)" (p.333)
52
Cannot use. (Roll Mode, Memory Divi-
sion)
If the waveform calculation function is used,
these functions cannot be used.
"10.1 Waveform Calcula-
tion Workflow" (p.224)
53
Cannot use. (Roll Mode, Memory Divi-
sion, Wave Calculation)
If one function is used, the other functions
cannot be used.
"7.2 Displaying Waveforms
During Recording (Roll
Mode)" (p.143)
54
Cannot use. (Overlay)
The overlay function is not available when us-
ing the Roll mode.
"7.3 Displaying New Wave-
forms Over Past Waveforms
(Overlay)" (p.144)
55
Cannot use. (Overlay, Memory Division,
Wave Calculation)
If roll mode is used, these functions cannot be
used.
"7.2 Displaying Waveforms
During Recording (Roll
Mode)" (p.143)
57
Cannot set. (External sampling)
Cannot use roll mode when using external
sampling.
"7.2 Displaying Waveforms
During Recording (Roll
Mode)" (p.143)
58
Rated capacity/rated output error.
Rated capacity/Rated output has exceeded
selected range. Input correct value.
59
Cannot use. (Roll Mode, Wave Calcula-
tion)
If the memory division function is used, these
functions cannot be used.
"11.1 Recording Settings"
(p.239)
60
No waveform data.
Acquire waveform data.
-
68
Recording length is too long.
Shorten the measurement record length.
The maximum record length that can be calcu-
lated is 2,000 div.
"3.4.3 Recording Length
(number of divisions)" (p.68)
91
LAN: Bad IP address.
Check the IP address.
95
LAN: Connection timed out.
Check the transmission setting.
97
LAN: Network error.
Check the connections between the instru-
ment and connection destination.
Displayed Warnings
Msg
No.
Message
Remedial Action
Reference
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......