7.9 Making Detailed Settings for Input Modules
162
Renew Data
(Data Refresh)
The data refresh rate can be set to Fast, Normal, or Slow.
The default setting is
[Normal]
. This allows stable measurement while removing
noise. For quicker response, select
[Fast]
, but note that this will make the mea-
surement more susceptible to noise. For further improved measurement stability,
select
[Slow]
.
Select
Selections
Description
Fast
Data are updated approximately every 1.2 ms.
Normal
Data are updated approximately every 100 ms. (default setting)
Slow
Data are updated approximately every 500 ms.