17.6.1 8966 Analog Unit
Temperature and humidity
range for guaranteed accuracy
23 ± 5°C (73.4 ± 9.0°F), 20 to 80% RH (when zero adjustment is executed 30 minutes after power
on)
Product warranty period
3
year
Period of guaranteed accuracy
1 year
No. of input channels
2 channels
Measurement range
5, 10, 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20 V/div
Measurement accuracy
± 0.5% f.s.(Filter 5 Hz On)
Temperature characteristic
± 0.06% f.s./°C
Frequency characteristic
DC Coupling: DC to 5 MHz -3 dB
AC Coupling: 7 Hz to 5 MHz -3 dB (low cut-off frequency 7 Hz ± 50%)
Noise
1.5 mVp-p (typ), 2 mVp-p (max) (sensitivity range, with input shorted)
Common mode rejection ratio
80 dB minimum (at 50/60 Hz and with signal source resistance 100
Ω
maximum)
Low-pass filter
OFF, 5 ± 50%, 50 ± 50%, 500 ± 50%, 5k ± 50%, 50 k ± 50%, 500 k ± 50%(Hz)-3dB
Input type
Unbalanced (floating)
Input coupling
AC/ DC/ GND
Input resistance
1 M
Ω
± 1%
Input capacitance
30 pF ± 10 pF (at 100 kHz)
A/D resolution
12 bit
Maximum sampling rate
20 MS/s
Input terminals
Insulated BNC terminal
Maximum input voltage
400 VDC
Maximum rated voltage to
earth
300 VAC/DC (between input channels and chassis, and between input channels),
Measurement category II (anticipated transient overvoltage 2500 V)
Operating temperature and
humidity
Same as the host Memory HiCorder
Operating environment
Same as the host Memory HiCorder
Storage temperature and hu-
midity
Temperature -10 to 50°C (14 to 122°F) Humidity 80% RH or less (non-condensating)
Dimensions
Approx. 106 mmW × 19.8 mmH × 207.5 mmD (4.17"W × 0.78"H × 8.17"D)
Mass
Approx. 250 g (8.8 oz)
Effect of radiated radio-fre-
quency electromagnetic field
± 15% f.s. (max) at 3 V/m
Effect of conducted radio- fre-
quency electromagnetic field
±45% f.s. (max) at 3 V (100 mV/div range, with 1 VDC input)
Applicable standards
Safety
EN61010
EMC
EN61326 Class A
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......