9.3 Judging Calculation Results
214
Set the judgment criteria (upper and lower threshold values) by which to judge numerical calculation
results. Judgment criteria can be set for every numerical calculation.
Waveform acquisition processing depends on the trigger mode setting (Single or
Repeat) and the criteria specified to stop measuring upon judgment (GO, NG or
GO & NG).
9.3 Judging Calculation Results
Make Calculation Settings
Start Measurement
Acquire Data
Calculate
Display Calculation Results
Calculation Judgments
Save
Stop Measurement
Stop Criteria
Trigger Mode
Stop criteria
not met
[Repeat]
[Single]
Measurement repeats
until stop criteria are met
Measurement repeats until you
click
[STOP]
Stop Criteria Met or
Stop Criteria is
GO&NG
Automatic saving will not be carried out before termination conditions are fulfilled
at the end of the calculation evaluation.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......