12.3 Setting FFT Analysis Conditions
247
11
Chapte
r 12 FF
T Function
12
About the frequency range and number of analysis points
___________________
• The settings for the frequency range and number of analysis points determine
the input signal acquisition time and frequency resolution.
• The frequency range setting for the FFT function corresponds to the timebase
setting of the Memory function. Changing the frequency range also changes
the data sampling period.
See:
"Relationship Between Frequency Range, Resolution and Number of Analysis
Points" (p.248)
• The cut-off frequency of the anti-aliasing filter is the same as the frequency
range setting.
See:
"Relationship Between Frequency Range, Resolution and Number of Analysis
Points" (p.248)
• The set number of analysis points specifies the amount of data to be analyzed
with each measurement. Increasing the number of analysis points increases
the frequency resolution, but also increases the time required for calculations.
See:
"Number of Analysis Points" (p.A15)
12.3.3 Setting the Frequency Range and Number of
Analysis Points
When the
[Reference]
is
[From Memory]
, the frequency is automatically
set.
The setting cannot be changed.
To control sampling by an external signal,
select
[External]
When the input data [Reference]
is
[From Memory]
The frequency range is set automatically
when analysis is started.
1
Set the number of FFT analysis points.
Move the flashing cursor to the
[Sampling Point]
item.
Select
See:
"Number of Analysis Points" (p.A15)
2
Select the frequency range.
Move the flashing cursor to the
[Frequency Range]
item.
Select
See:
"Relationship Between Frequency Range, Resolution and Number of
Analysis Points" (p.248)
1000 (default setting), 2000, 5000, 10000
8 MHz(default setting), 4 MHz, 2 MHz, 800 kHz, 400 kHz, 200 kHz, 80
kHz, 40 kHz, 20 kHz, 8 kHz, 4 kHz , 2 kHz, 800 Hz, 400 Hz, 200 Hz, 80
Hz, 40 Hz, 20 Hz, 8 Hz, 4 Hz, 1.33 Hz, 800 mHz, 667 mHz, 400 mHz, 333
mHz, 133 mHz, External
Procedure
To open the screen: Right-click and select
[STATUS]
→
[Status]
sheet
See:
To set from the Waveform screen (p.261)
Frequency Resolution
(during acquisition)
The resolution is affected by settings of
frequency range and the number of
analysis points. Not displayed for exter-
nal sampling.
2
1
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......