7.9 Making Detailed Settings for Input Modules
172
Channels installed with Model MR8790 cannot be measured.
See:
"Opening the [Each Ch] sheet, Making a Channel Selection" (p.159)
Type
Selects the waveform type.
Select
Frequency
Sets the frequency of output signal.
Amplitude
Sets the amplitude of output signal.
The output voltage with guaranteed accuracy is the sum of the amplitude and the
offset, between -10 V and +10 V. If the sum of the amplitude and the offset is set
outside the guaranteed accuracy range, parts of the waveform will be clamped to
the upper limit, approxi14 V and the lower limit, approximately -14 V.
Offset
For DC output: Sets DC voltage.
For sine wave output: Sets the offset voltage.
The output voltage with guaranteed accuracy is the sum of the amplitude and the
offset, between -10 V and +10 V. If the sum of the amplitude and the offset is set
outside the guaranteed accuracy range, parts of the waveform will be clamped to
the upper limit, approxi14 V and the lower limit, approximately -14 V.
7.9.10 MR8790 Waveform Generator Unit Settings
Selections
Description
DC
DC output (Default setting)
Sine
Sine wave output
0 Hz to 20000 Hz
0.000 V p-p to 20.000 V p-p
-10 V to +10 V
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......