7.4 Converting Input Values (Scaling Function)
152
1. For scaling, set the
[Method]
to
[Ratio]
.
2. Point the flashing cursor to the conversion rate setting and select
[dB Scal-
ing]
.
3. In the displayed entry field and physical quantity field, enter 40 dB and 60 dB,
respectively.
After entering the values, select
[Confirm]
.
See:
"7.1.3 Alphanumeric Input" (p.139)
The conversion rate for the entered dB value is entered. (The offset becomes 0.)
Using the dB value
Example 5
Acquiring the conversion rate to convert 40 dB input to 60 dB
1
2
2