![Hioki MR8740 Скачать руководство пользователя страница 222](http://html1.mh-extra.com/html/hioki/mr8740/mr8740_instruction-manual_2137324222.webp)
9.2 Settings for Numerical Value Calculation
210
9.2 Settings for Numerical Value Calculation
1
2
Procedure
To open the screen: Right-click and select
[STATUS]
→
[Num Calc]
sheet
1
Enable the Numerical Calculation function.
Move the flashing cursor to the
[Numerical Calc]
item.
Select
[On]
.
2
Specify the Numerical Calculation range.
Move the flashing cursor to the
[Calc Area]
item.
Select
See:
"6.2 Specifying a Waveform Range (A/B Cursor)" (p.122)
3
Perform calculation settings.
(1)
Select the calculation type.
Move the flashing cursor to the number of the calculation type
for which to make settings, and select the calculation type.
(When you select
[List]
a list of calculation types appears.)
WholeWave
Applies calculations to the whole
waveform. (default setting)
A-B Wave
Applies calculations to the data be-
tween A/B cursors.
AfterTrigger
Applies calculations to the waveform
after a trigger.
.Calculation No.
Calculation
Type
(1)
(2)
(3)
Parameters*
*: Setting choices depend
on the calculation type.
Channel for
Calculation
(4)
Statistical
calculation
(5)
Phenomenon
*2: Setting can also be made for logic channels
X-Y Area*
1
Area of X-Y composite waveform
Time to Lev*
2
Time from trigger to specified level
Lev-time
Measurement value at a specified time
point after triggering
Pulse Width*
2
Pulse width of waveform data
Duty Ratio*
2
Duty of waveform data
Pulses*
2
Pulse count of waveform data
Calculation
Four arithmetic operations on numerical
calculation results
Time Diff*
2
Time difference between phenomenon A
and B.
Phase Diff*
2
Time difference between phenomenon A
and B displayed as a phase contrast.
High Level*
1
High level value for waveform data
Low Level*
1
Low level value for waveform data
*1: Calculation performed with MR8990 Digital Volt-
meter Unit is invalid. The indication of the calcu-
lation result becomes "******".
Off
No calculation. (default setting)
Average
Average value of waveform data
Rms*
1
RMS value of waveform data
Peak-Peak
Peak-to-peak value of waveform data
Maximum
Maximum value of waveform data
Max-Time
Time from trigger to maximum value
Minimum
Minimum value of waveform data
Min-Time
Time from trigger to minimum value
Period*
2
Period of signal waveform
Frequency*
2
Frequency of signal waveform
RiseTime*
1
Rise time of waveform data
FallTime*
1
Fall time of waveform data
Std Dev*
1
Standard deviation of waveform data
Area*
1
Area enclosed by zero position and
signal waveform
When selecting
[A-B Wave]
, specify the
calculation range using the A/B cursors on
the Waveform screen.
If no measurement data has been ac-
quired by the instrument, first measure
once so that the range can be specified for
calculations to be applied to subsequent
measurements.
Also select
[Judge]
if you require judg-
ment of calculation results.(p.214)
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......