10.2 Settings for Waveform Calculation
231
9
Chap
ter 10
W
avefo
rm Calculation
Functions
10
10.2.3 Changing the display method for calculated
waveforms
Procedure
To open the screen: Right-click and select
[STATUS]
→
[Wave Calc]
sheet
1
Enable waveform display, and display color
Move the flashing cursor to the
[Wave]
column.
Select
2
Select a method to set scaling
Move the flashing cursor to the
[Scale]
column for the Calculation No.
to be set.
Select
3
Set the upper and lower limits of the display range (when
[Manual]
is selected)
Select
[Lower]
and
[Upper]
.
Select an entry method and enter the limit values.
Entry range: -9.9999E+29 to +9.9999E+29
See:
"7.1.3 Alphanumeric Input" (p.139)
4
Specify the physical units
Move the flashing cursor to the
[Unit]
column.
Select an entry method and enter the physical units.
See:
"7.1.3 Alphanumeric Input" (p.139)
5
Select the graph to be displayed.
(When split screen (
[Format]
item on the
[Status]
sheet) is
[Dual]
or higher)
Move the flashing cursor to the
[Graph]
column and select the graph
b
Waveform
color
Upper and
lower limits
Displayed
measurement units
Display range
setting method
Calculation No.
To copy settings
between Calculation Nos.:
Click the calculation No.
1 2
4
3
Graph to
display
6
On-Off
Set On to display the waveform of the flashing cursor
column (default setting). Set to Off to hide display.
↑↓
Select the waveform color.
All On-Off
Select On to display all waveforms. Select Off to hide all
waveforms.
Auto
Automatically sets the display range of the vertical axis. (After
calculation, the upper and lower limits are obtained from the
results, and set automatically.)
Manual
Upper and lower limits of the vertical axis display range are
entered manually.
A shorter calculation time than with Auto is possible.
Depending on calculation results, auto-
matic scaling settings may be unsatisfac-
tory, in which case the limits must be
entered manually.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......