3.5 Input Channel Setting
77
3
Chapte
r 3 Me
asu
remen
t Proced
ure
Make settings for the logic channels. The channel settings window (Logic sheet) is shown when the
display format is 1, 2, 4, 8 or 16 screens.
1. Logic Width
Allows changing the display width of the logic waveform.
Making waveforms more narrow can enhance the readability of the display when
there are a high number of waveforms.
Select
2. Waveform
Display
Position
Determines where on the screen the logic waveform is displayed.
The position can be freely moved within the range of the display.
3. Waveform
Display Color
Specifies the color in which the waveform of the selected channel is displayed.
You can also select the same color as another channel.
For logic modules, the color can be specified for each module and each channel
separately.
Select
3.5.3 Logic Channel
1.
Procedure
To open the screen: Right-click and select
[DISP]
→
Waveform screen
→
Right-click and select
[CH.SET]
→
Channel settings window (
[Logic]
sheet)
2.
3.
1
Move the flashing cursor to the channel for
which to make settings.
2
Select the settings by clicking the mouse.
To copy the settings of one channel to another
See:
"7.8 Copying settings to other channels (calculation No.) (Copy function)" (p.158)
Wide
Make the waveform wider.
Normal
Display the waveform at normal width.
Narrow
Make the waveform more narrow. (default setting)
Off
The waveform is not displayed. If the
[Save Channel]
setting is
[Disp Ch]
, data
for the channel will not be automatically saved.
See: "4.2.2 Automatically Saving Waveforms" (p.86)
On
The waveform is displayed. Set the display color by clicking
[
↑
]
or
[
↓
]
.
Probe On-Off
Switches the waveform display of the same probes to all ON or all OFF.
All On-Off
Switches the display of all logic waveforms to all ON or all OFF.
This can be selected when the cursor at in the waveform display position item.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......