7.6 Fine Adjustment of Input Values (Vernier Function)
156
Fine adjustment of input voltage can be performed arbitrarily on the Waveform screen. When
recording physical values such as noise, temperature and acceleration using sensors, amplitude
can be adjusted to facilitate calibration.
7.6 Fine Adjustment of Input Values (Vernier
Function)
Normal Display
Vernier Function Enabled
1.2 V input is displayed as 1.0 V
1.2 V
1.0 V
Procedure
To open the screen: Right-click and select
[DISP]
→
Waveform screen
→
Right-click and select
[CH.SET]
→
Channel settings window (
[Analog]
sheet)
1
Move the flashing cursor to the
[ ]
vernier setting item of
the channel to adjust.
2
Make the adjustment while watching the waveform.
Vernier
Magnify waveform
Vernier
Compress waveform
Vernier Reset
Return waveform to original position
• The adjustment range is 50 - 200% of the original waveform. The magnifica-
tion/compression ratio is not displayed.
• Vernier adjustments cannot be verified on waveforms.
• The waveform data (saved file data) is adjusted by the Vernier function.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......