Operating Precautions
10
Before Connecting Cables
When measuring power line voltage
• Connecting cables should only be connected to the secondary side of a
breaker Even if there is a short circuit on the secondary side of the breaker, the
breaker cuts off the electric supply. Do not connect to the primary side of a
breaker because unrestricted current flow could damage the instrument and
facilities if a short circuit occurs.
• To prevent electrical shocks and personal injury, do not touch any input termi-
nals on the VT (PT), CT or the instrument when they are in operation.
• Do not permanently connect the instrument in an environment where voltage
surges exceeding the maximum input voltage may occur. Applying voltage
may result in damage to the instrument, or a serious accident.
• Do not short-circuit two wires to be measured by bringing the connection
cables into contact with them. Arcs or such grave accidents are likely to occur.
• To avoid short circuit or electric shock, do not touch the metal parts of the con-
necting cable clips.
• To avoid electrical shocks, be careful to avoid shorting live lines with the con-
nection cable clips.
• To avoid electric shock and short-circuit accidents, use only the specified test
leads to connect the instrument input terminals to the circuit to be tested.
• To prevent an electric shock, confirm that the white or red portion (insulation
layer) inside the cable is not exposed. If a color inside the cable is exposed, do
not use the cable.
• The cable is hardened under the 0°C (32°F) or colder environment. Do not
bend or pull cables in such environments to avoid tearing insulation or break-
ing the cable.
•
Connecting to the BNC jacks on modules
Do not use a metal BNC connector. If you connect a metal BNC cable to an
insulated
BNC connector, the insulated BNC connector can be damaged and the instru-
ment may be damaged.
• To prevent cable damage, do not step on the cable or pinch them between
other objects. Do not bend or pull cables at their base.
• Use only the specified connection cables. Using a non-specified cable may
satisfy the specification requirements due to poor connection or other reasons.
• For detailed precautions and instructions regarding connections, refer to the
instruction manuals for your modules, connection cables, etc.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......