3.5 Input Channel Setting
78
The input channel settings can be set differently for each displayed sheet. Up to four sheets can be set.
You can set desired waveforms for displaying to different sheets and switch them.
• When the standard logic display is on, the 8970 Freq Unit installed on unit 1 or 2 can no
longer be used. Furthermore, the 16-bit resolution 8967 TEMP Unit, 8968 High Resolu-
tion Unit, 8969 Strain Unit, U8969 Strain Unit and U8974 High Voltage Unit have a reso-
lution of 12 bits. Also, when MR8990 Digital Voltmeter Unit is installed on unit 1 (unit 1
or unit 2 in the case of MR8741), the standard logic can no longer be used.
• With MR8740, install the logic units on unit 1 to unit 8. Even if you install logic units on
units 9 and after, they will be invalid.
3.5.4 Display Sheet
Sheet switching (1 to 4)
• Only the following display-related settings can be set to each displayed sheet.
Analog waveform:
Display ON/OFF, waveform color, ratio, zero position,
vernier, invert, graph variable (ON/OFF, upper and
lower limit)
Logic waveform:
Display ON/OFF, waveform color, display position, and
logic width X-Y waveform: Display ON/OFF, waveform
color, X ch/Y ch waveform calculation (X ch, Y ch)
Waveform calculation: Display ON/OFF, waveform color, scale setting, upper/
lower limit, graph
Common setting:
Display format
• The measurement-related settings other than above will be common to all dis-
played sheets.
When range is changed, the range of all display sheets is changed.
• When a settings file is saved, the settings of all displayed sheets are saved.
• A waveform file is saved based on the setting of the sheet displayed at the
time of saving. When a waveform file is loaded, only the sheet displayed at the
time of saving can be loaded because other sheets were not saved.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......