3.5 Input Channel Setting
74
Set the analog channel.
For information on specific settings for 8967 TEMP Unit, 8969 Strain Unit, U8969 Strain Unit, 8972
DC/RMS Unit and MR8990 Digital Voltmeter Unit, and U8974 High Voltage Unit see "7.9"(p.159).
1. Waveform
Display Color
Specifies the color in which the waveform of the selected channel is displayed.
You can also select the same color as another channel.
Select
2. Vertical axis
(Voltage axis)
Range
Sets the vertical axis (voltage axis) range for each channel. The value set here is
the voltage of one increment on the vertical axis.
For information on the full-scale value for the respective modules, see the table
in the section "6. Zero Position" (p.75).
When the Variable function is on but Variable Auto Adjustment is off, the size of
the waveform on the screen will not change even if the vertical axis (voltage)
range is changed.
3. Coupling
Set the input signal coupling method. Normally, DC coupling should be selected.
Select
3.5.2 Analog Channel
Procedure
To open the screen: Right-click and select
[DISP]
→
Waveform screen
→
Right-click and select
[CH.SET]
→
Channel settings window (
[Analog]
sheet)
1.
2.
3. 4.
5.
6.
7.
1
Move the flashing cursor to the channel for
which to make settings.
2
Select the settings by clicking the mouse.
To copy the settings of one channel to another
See:
"7.8 Copying settings to other channels (calculation No.) (Copy function)" (p.158)
Off
The waveform is not displayed. If the Auto Save setting for
[Save Channel]
is
[Disp Ch]
, data for the channel will also not be saved.
See: "Select the channel to save." (p.88)
On
The waveform is displayed. Set the display color by clicking
[
↑
]
or
[
↓
]
.
All On-Off
Switches the waveform display of all channels to all ON or all OFF
An over-range condition has occurred
Change the vertical axis (voltage axis) range to a lower sensitivity set-
ting.
DC (V)
Both DC components and AC components of the input signal will be measured.
(default setting)
AC ( )
Only AC components of the input signal will be measured. DC components are
blocked.
GND ( )
Input is shorted to ground. (Allows checking the zero position.)
V
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......