7.9 Making Detailed Settings for Input Modules
168
See:
Opening the
[Each Ch]
sheet, Making a Channel Selection (p.159)
Mode
Switches between voltage measurement and RMS measurement.
Select
Response
Response can be set to three speeds: Fast, Normal and Slow.
Normally set to
[Fast]
, this can be changed to
[Normal]
or
[Slow]
to stabilize the
display when measuring low frequencies, or when severe fluctuations are
present.
Select
7.9.7 Settings for the 8972 DC/RMS Unit
Selections
Description
DC
Voltage measurement (default setting)
RMS
RMS measurement
Selections
Description
Fast
Sets the response time to about 100 ms. (default setting)
Normal
Sets the response time to about 800 ms.
Slow
Sets the response time to about 5 s.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......