4.2 Saving Data
86
Measurement data are acquired for the recording length and then saved automatically each time.
Save location and data type are selected before the measurement.
Waveform data can be saved.
4.2.2 Automatically Saving Waveforms
Procedure
To open the screen: Right-click and select
[SYSTEM]
→
[File Save]
sheet
1
Enable auto save.
Move the flashing cursor to the
[Auto Save]
item, and select
[On]
.
Default setting: Off (automatic saving is not performed)
2
Set the save type.
Move the flashing cursor to the
[Save Type]
item.
Select
3
Set the save destination.
Move the flashing cursor to the
[Save To]
, and select
[Edit]
.
The Browse folders dialog box appears (at bottom right).
Move the flashing cursor to the save target media
*
and confirm
the setting with
[Confirm]
.
Select
When the root directory (topmost folder on media) was selected, a folder named
"
HIOKI_MR8740
" or "
HIOKI_MR8741
" is created automatically. (If the media was
initialized in the instrument, the folder will already have been created.) This folder
is then used as save target.
To create a new folder, select
[New Folder]
. When
[LAN]
is set
as the save destination,
[New Folder]
setting is ignored and the
folder named with the current date is created.
4
Set the file name.
Move the flashing cursor to the
[Save Name]
, and enter the save
name.
See:
"7.1.3 Alphanumeric Input" (p.139)
When
[LAN]
is set as the save destination,
[Save Name]
setting is
ignored and the files to be saved are named in the previously-
determined format.
See:
"Save Operations (When setting the save destination to [LAN])"(p.90)
Wave
Binary
Save the waveform data in binary format.
(Data saved in binary format can only be loaded to this instrument.)
Wave
Text
Save the waveform data in text format.
The data can be thinned out and then saved.
(This data can be opened by a computer editor or spreadsheet
software, but it cannot be loaded to this unit.)
USB1
Automatically save the waveform data on the USB memory stick.
LAN
Automatically save the waveform data on the PC connected via
LAN. Model 9333 LAN Communicator is required to be installed.
1
2
3
4
File name
The maximum number of characters for the
[Save Name]
string is 123. The
maximum path length including file name is 255 characters.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......