7.9 Making Detailed Settings for Input Modules
163
6
Chapte
r 7 Utility Functions
7
The 8969 Strain Unit and U8969 Strain Unit can perform auto balance.
When auto balance is performed, the reference output level of the conversion
unit can be matched with the specified zero position.
It is applicable only to a 8969 Strain Unit and U8969 Strain Unit.
Before executing auto-balance ___________________________________
• Turn power on and wait 30 minutes to allow the internal temperature of the
input module to stabilize.
• With the strain gauge transducer connected to the input module, execute
auto-balance under stable input conditions.
• Auto-balance cannot execute during measurement.
• Mouse operations are not accepted while auto-balance is executing.
To execute auto-balance_________________________________________
See:
Opening the
[Each Ch]
sheet, Making a Channel Selection (p.159)
Auto Balance can also be executed from the Channel settings window (Analog
sheet) (if the range of a channel with installed strain unit is selected).
See:
Opening the Channel settings window (
[Analog]
sheet): (p.74)
In the following cases, auto-balance should be executed again.
• After changing the vertical axis (strain axis) range
• After an input module has been removed or inserted
• After the strain gauge transducer has been replaced
• After power has been turned off and on
• After performing a system reset
• When ambient temperature has changed significantly (the zero position may
drift)
7.9.4 Settings for the 8969 and U8969 Strain Unit
The instrument describes Model U8969 as “8969”.
Setting Item:
[Range]
Selections
Description
Auto Bal All
Chs
Auto-balance will be executed for all channels where a Model 8969
or U8969 Strain Unit is installed.
Auto Bal Ch 1
Auto-balance will be executed for the currently selected channel.
If "Warning: Auto balance failed." appears:
The channel on which auto-balance failed is displayed.
Verify the following, and execute again:
• Is the strain gauge transducer in a discharged state? (Make sure that it is not
being subject to vibration, etc.)
• Is the strain gauge transducer connected correctly?
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......