![Hioki MR8740 Скачать руководство пользователя страница 173](http://html1.mh-extra.com/html/hioki/mr8740/mr8740_instruction-manual_2137324173.webp)
7.9 Making Detailed Settings for Input Modules
161
6
Chapte
r 7 Utility Functions
7
See:
Opening the
[Each Ch]
sheet, Making a Channel Selection (p.159)
Mode
Set to match the type of thermocouple being used.
Select
RJC (Reference
Junction Compen-
sation)
When connecting a thermocouple directly to the input module, select
[Int]
.
Reference junction compensation is performed within the input module. When
connecting through a reference junction device (e.g., a 0°C control tank), select
[Ext]
.
Select
Burn Out
A broken thermocouple wire can be detected during temperature measurement.
Normally when a thermocouple wire breaks, measured values exhibit random
instability.
Select
7.9.3 Settings for the 8967 TEMP Unit
Selections
Measurement Range
Selections
Measurement Range
Temp- K
-200 to 1350°C
Temp- R
0 to 1700°C
Temp- J
-200 to 1100°C
Temp- S
0 to 1700°C
Temp- E
-200 to 800°C
Temp- B
400 to 1800°C
Temp- T
-200 to 400°C
Temp- W
0 to 2000°C
Temp- N
-200 to 1300°C
Selections
Description
INT
Reference junction compensation is provided within the input mod-
ule. (default setting)
(Measurement Accuracy: The sum of the accuracies of the temper-
ature measurement and the reference junction compensation.)
EXT
Reference junction compensation is not provided within the input
module.
(Measurement Accuracy: The accuracy of temperature measure-
ment only)
Selections
Description
Off
Broken wires are not detected.
On
Broken wires are detected.
Wire breakage is detected by sensing a miniscule current flow
(about 100 nA) through the thermocouple. If the thermocouple
wires are long or composed of a high-resistance material, set
[Burn
Out]
to
[Off]
to avoid measurement errors.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...