7.9 Making Detailed Settings for Input Modules
171
6
Chapte
r 7 Utility Functions
7
See:
"Opening the [Each Ch] sheet, Making a Channel Selection" (p.159)
Mode
Switches between voltage measurement and RMS measurement.
Select
Response
The response time for RMS measurement can be set to three speeds: Fast, Nor-
mal and Slow.
The response time can be set to
[Slow]
to stabilize the measured value when
the frequency is low or when severe fluctuations are present.
Select
7.9.9 Model U8974 High Voltage Unit Settings
Selections
Description
DC
Voltage measurement (Default setting)
RMS
RMS measurement
Selections
Description
Fast
Response time is set to 150 ms.
Normal
Response time is set to 500 ms. (Default setting)
Slow
Response time is set to 2.5 s.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......