212
External Input and Output
3
Click the
[OUT1]
and
[OUT2]
boxes in the
[External out]
area in turn, and then from the list,
choose a signal output action.
Choose a condition where the instrument outputs a signal.
Judge(Pass)
Outputs a low-level signal when a pass judgment is given for the numerical calculation
result.
Judge(Fail)
Outputs a low-level signal when a fail judgment is given for the numerical calculation
result.
Error
Outputs a low-level signal when an error occurs.
Busy
Outputs a low-level signal while rejecting a START signal because the instrument is in
the busy state such as performing a measurement and saving data.
Waiting trigger
Outputs a low-level signal while waiting for a trigger.
The instrument continuously outputs the signal for a pass/fail judgment (low-level output) until it starts the next
measurement.
The instrument can output various signals depending on its state.
Output signal
Open-drain output (with voltage output), active low
Output voltage
range
High level:
4.0 V to 5.0 V
Low level:
0 V to 0.5 V
Maximum input
voltage
50 V DC, 50 mA, 200 mW
GND
OUT1
OUT2
100
kΩ
10
kΩ
5 V
High
4.0 V to 5.0 V
Low
0 V to 0.5 V
Output period
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......
Страница 238: ......