58
Configuring Measuring-Module-Specific Settings
5
Click the
[
Calibration]
box, and then choose a calibration setting from the list.
Enabling this setting automatically calibrates the module or synchronizes the channels at the start of
measurement. The synchronization between channels allows integration calculations to synchronize with each
other.
Off
Does not calibrate the module nor synchronize the channels.
On
Calibrates the instrument and synchronizes the channels.
Synchronization
Synchronizes between channels only.
•
It takes about 150 ms to calibrate the module. During this period of time, no measurement is performed.
•
If the channels are synchronized with each other, the instrument sends a signal that interrupts the integration
to each module at the start of measurement; thus, the instrument has to wait until the first integration
finishes.
The wait time required for this process stands at the sum of 10 ms and integration time*.
*: The integration time varies depending on the NPLC setting.
Even when synchronization is not performed, the wait time described above is required for measurement
performed immediately after the settings of Model MR8990 Digital Voltmeter Unit has been changed. No
wait time is required when the measurement is performed without any setting changes
•
When
[Calibration]
is set to
[Off]
(default setting), execute calibration manually.
Refer to “2.9 Executing Calibration (For the Instrument With Model MR8990 Installed)” in Quick Start
Manual.
6
Click
[Close]
.
The setting dialog box closes.
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......
Страница 238: ......