128
Searching For Positions Where a Trigger Condition Is Satisfied
(2) With the
[In]
or
[Out]
setting
[In]
Upper limit value
Lower limit value
[Out]
Upper limit value
Lower limit value
Setting
Description
Filter
Off
,
10
to
10,000
Allows you to enter a filter width in the number of samples. Positions
where a search criterion is continuously satisfied during the specified
period are determined to be the retrieved positions.
Upper limit
value
−
3.4028
E+
38
to
3.4028
E+
38
Default:
0.2
Allows you to enter an upper limit value.
Lower limit
value
−
3.4028
E+
38
to
3.4028
E+
38
Default:
0.2
Allows you to enter a lower limit value.
-2. When a logic channel is selected under
[Target channel]
You can search for a position that matches the specified pattern.
Trigger pattern
LA1 1
LA2 0
LA3 X
LA4 X
Setting
Description
Satisfied
OR
The positions where any one of the specified search patterns is
satisfied are determined to be the retrieved positions.
AND
The positions where all the specified patterns are satisfied are
determined to be the retrieved positions.
Filter
Off
,
10
to
10,000
Allows you to enter a filter width in the number of samples.
Positions where a search condition is continuously satisfied
during the specified period are determined to be the retrieved
positions.
Trigger pattern
X
Ignored data.
0
Searches for low-level positions.
1
Searches for high-level positions.
In the logic search, the retrieved position is the point where the condition that has not been satisfied changes
to being satisfied. Thus, even if signals match a search pattern at the start of a search, it is not regarded as a
retrieved position.
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......