Configuring the Input Channel settings
4
Configure the input coupling, low-pass filter, and probe ratio settings.
Click the area that includes
[Coupling]
allows the setting dialog box to appear.
(1)
(2)
(3)
(1) Click the
[
Coupling]
box, and then choose a coupling method for an input signal from the
list.
Choose a coupling method for an input signal. In general, use the DC coupling.
DC
Measures both DC and AC components of an input signal.
AC
Measures an AC component only of an input signal. A DC component can be
eliminated.
GND
Connects the input terminal to the ground, which allows you to check the zero position.
(2) (1) Click the
[L.P.F]
box, and then choose a cutoff frequency of the low-pass filter from the
list.
Enabling the low-pass filter installed in the module can eliminate unwanted high-frequency components.
The filters available vary depending on the module types. Use an adequate filter in accordance with the
characteristics of an input signal.
Example: Model 8966 Analog Unit
OFF
,
5 Hz
,
50 Hz
,
500 Hz
,
5 kHz
,
50 kHz
,
500 kHz
(3) Click the
[
Probe ratio]
box, and then choose a probe ratio from the list.
Choose any of the ratios when the measurement involves use of a connection cord or probe.
1:1
Choose this ratio when using Model L9197, Model L9198, Model L9790, or Model
L9217 Connection Cord.
1:10
Choose this ratio when using Model 9665 10:1 Probe.
1:100
Choose this ratio when using Model 9666 100:1 Probe, Model P9000-01 Differential
Probe, or Model P9000-02 Differential Probe.
1:1000
Choose this ratio when using Model 9322, Model P9000-01, or Model P9000-02
Differential Probe.
1
Measurement Method
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......
Страница 238: ......