59
Configuring Measuring-Module-Specific Settings
Configuring Model U8974 High Voltage Unit settings
> [Channel] > [U8974]
3
1
2
1
In the
[Mode]
area, click the
[DC]
or
[RMS]
to choose a measurement mode.
DC
For voltage measurement
RMS
For RMS measurement
2
(When you have changed the measurement mode) Click
[Zero adjust]
.
The instrument performs zero-adjustment. Execute zero-adjustment without any input.
3
Click the area that includes
[Response]
.
The setting dialog box appears.
4
Click the
[Response
]
box, and then from the list, choose a response time for RMS
measurement.
Fast
Sets the response time at 150 ms.
Normal
Sets the response time at 500 ms.
Slow
Sets the response time at 2.5 seconds.
Setting the response time to
[Slow]
can stabilize measured values if frequency is relatively low or voltage
fluctuates severely.
5
Click
[Close]
.
The setting dialog box closes.
3
Advanced Functions
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......
Страница 238: ......