Setting the Trigger Satisfaction Conditions (AND/OR Operation) Among Trigger Sources
5.5
Setting the Trigger Satisfaction Conditions
(AND/OR Operation) Among Trigger Sources
Set the trigger logical-conditions among the analog, logic, external, and interval triggers by
choosing between logical AND or OR operation.
The forcible trigger triggers the instrument regardless of the trigger logical-conditions setting. If all
trigger sources are set to off (i.e., with no trigger setting), recording starts immediately (free run).
> [Trigger] > [Common]
Click
[AND]
or
[OR]
, whichever is displayed, to switch to the other.
OR
When any one of the specified trigger conditions changes from not being satisfied to
being satisfied (at a changing point), the instrument is triggered. Thus, even though
a trigger condition has been already satisfied at the start of waiting for a trigger, the
instrument is not triggered until a changing point is detected.
AND
Only when all of the specified trigger conditions are satisfied, the instrument is
triggered. Thus, if all the specified trigger conditions have been already satisfied at the
start of waiting for a trigger, the instrument is immediately triggered.
Setting example: To trigger the instrument when a waveform crosses the zero-volt level in
the positive direction (
[ ]
*)
The instrument is triggered based on whether the trigger satisfaction condition is set to AND or OR
operation in the following ways:
Channel
Trigger
Level at time
Slope
Filter
CH1, CH2
Level
0.00 V
*
Off
[AND]
[OR]
One waveform is above 0 V,
and the other is also above 0 V.
Either waveform crosses 0 V
upwards.
Start
Start
*: With the
[AND]
setting, the slope setting item appears as
[HIGH]
.
5
Configuring the
Trigger Settings
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......
Страница 238: ......