Configuring Generator-Module-Specific Settings
Comprehensive test
The instrument outputs the following voltages and currents of each output function and each
generation range, measuring them with the internal measurement circuit. Then, the instrument
sends back measured values of each output value.
Output item
Generation
range
Measurement
range
Output testing
point
Voltage
5 V
0.5 V
0 V
0.5 V
5 V
0 V
5 V
Current
5 mA
5 mA
5 mA
1.1 mA
1 mA
1 mA
1 mA
275 µA
250 µA
250 µA
250 µA
55 µA
50 µA
50 µA
50 µA
0 µA
CAUTION
Do not connect anything with the OUTPUT terminal during comprehensive tests. The
instrument outputs a zero-volt signal and connects the OUTPUT terminal with the GND
terminal (in the short-circuit state) during comprehensive tests. If a power supply device
connects with the instrument, a short-circuit current will flow, resulting in damage to the
instrument or power supply device. The comprehensive test results may be affected
depending on devices connected.
Offset canceling function
Regulating the zero point of the output circuit
The instrument measures an output circuit offset with the internal measurement circuit and outputs
a value the offset value is subtracted from.
CAUTION
Do not connect anything with the OUTPUT terminal during an offset measurement. The
instrument outputs a zero-volt signal and connects the OUTPUT terminal with the GND
terminal (in the short-circuit state) during the offset measurement. If a power supply
device connects with the instrument, a short-circuit current will flow, resulting in damage
to the instrument or power supply device. The results (zero-point regulation) may be
affected depending on devices connected.
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......
Страница 238: ......