![Hioki MR8740-50 Скачать руководство пользователя страница 216](http://html1.mh-extra.com/html/hioki/mr8740-50/mr8740-50_instruction-manual_2137326216.webp)
210
External Input and Output
12.1
External Input and Output
External input (IN1), (IN2)
Externally inputting signals can start and stop measurement as well as save data.
In factory default settings, the START signal is assigned to the IN1, and the STOP signal to the IN2
terminal.
How to input signals
1
Connect each of the IN1, IN2, and GND terminals to an external signal-outputting device with
wires.
Refer to “2.3 Connecting the External Control Terminals” in Quick Start Manual.
2
> [System] > [External terminal]
3
3
Click the
[IN1]
and
[IN2]
boxes in the
[External In]
area in turn, and then from the list, choose
an action performed when an input signal is inputted.
START
Starts a measurement.
STOP
Stops the measurement (post-measurement processes such as numerical calculations
and automatic saving are performed).
START/STOP
Starts a measurement when the low-level signal is inputted; stops the measurement
when the high-level signal is inputted.
SAVE
Saves data on the specified conditions into a storage device that is specified in (
[Status]
> [Save] > [Save Icon operation] >
)
[Quick]
.
ABORT
Forcibly terminates the measurement. (No post-measurement processes such as
numerical calculations or automatic saving are performed. However, the instrument
automatically saves a waveform file that contains data acquired until the forcible
termination)
EVENT
Puts on an event mark.
The event marks are put on the waveform screen.
You can put event marks by clicking the start icon during a measurement.
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......
Страница 238: ......