69
Configuring Generator-Module-Specific Settings
Configuring Model MR8791 Pulse Generator Unit settings
1
Click
[MR8791]
.
6
1
3
4
5
2
2
Click the
[Mode]
box, and then choose an output type from the list.
PULSE
Outputs a pulse waveform.
PATTERN
Outputs a pattern waveform.
3
Configure the pulse output settings.
Frequency
Allows you to enter a pulse output frequency. (In 0.1 Hz increments)
0.0
Hz to
20000.0
Hz
Duty ratio
Allows you to enter a pulse duty ratio. (In 0.1 percent point increments)
•
With the 0% setting, the low-level signal is outputted; with the 100% setting, the high-level signal is
outputted. No pulse is outputted.
•
With the 100% setting, the low-level signal is outputted even if the output setting is disabled.
•
The output pulse width setting of 1 µs, which can be derived from the frequency and duty ratio settings, may
cause absence of pulses.
0.0
% to
100.0
%
3
Advanced Functions
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......
Страница 238: ......